PCR cleanup by Gibco Concert kit
Note: We elute with Tris-HCl rather than the TE provided
by the manufcturer
-
Preheat an aliquot of 10mM Tris-HCl pH 8.0 on the 65oC
block
-
Add 400 uL of Binding Solution H1 to 50 uL PCR reaction and
mix.
-
Put spin cartridge into 2 mL collection tubes provided and
spin the tube at 12000 x g for 1 minute
-
Discard flow-through
-
Add to each column 700 uL of wash buffer H2 (make sure ethanol
has been added)
-
Spin the tube at 12000 x g for 1 minute
-
Discard flow-through and again spin the tube at 12000
x g for 1 minute
-
Transfer each column to a fresh 1.5 mL Eppendorf tube
-
Add 50 uL 10 mM Tris-HCl, pH 8.0 directly to the center of
the spin column. Incubate at room temperature for 1 min then spin at 12000
x g for 2 min. (Sometimes sterile water is used instead of Tris-HCl)